This electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
- Insulation and AC/DC continuous voltage withstand testing
- Safety standards compliance testing
- Stores up to 10 sets of test conditions (10 each for insulation and voltage withstand)
- Zero-V Switching (voltage withstand testing)
- Electrical Discharge function (DC voltage withstand, insulation testing)
- Automated testing support
- Interlock Protect function
- Remote controllable
- Programable testing, full remote control, Automatic Insulation Withstanding Tester
Basic specifications : Withstanding test section
| Testing voltage | 0.2 kV to 5.00 kV AC, 500VA (maximum 30 minutes) 0.2 kV to 5.00 kV DC, 50VA (continuance) |
| Voltage setting method | Digital setting |
| Waveform | Same as the power supply waveform |
| Frequency | Same as the power supply frequency |
| Measurement range | Current: 0.01 to 100.0 mA, ±(2 % f.s. +5 dgt.) 10 mA/100 mA AC (Average value rectified, effective value display) |
Basic specifications : Insulation test section
| Testing voltage | 50 V to 1000 V DC |
| Measurement range | 0.1 m-ohm to 9999 m-ohm, 4 ranges |
Basic specifications : general
| Judgment function | Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method) |
| Timer section | Setting range: 0.3 to 999 seconds |
| Interfaces | EXT I/O, EXT SW, RS-232C, GP-IB |
| Display | Fluorescent tube display (digital), Analog meter |
| Monitor function | Output voltage, detection current |
| Power supply | 100 V to 120 V AC / 200 to240 V AC, (50/60 Hz), 1000 VA max. |
| Dimensions, mass | 320 mm(12.60 in)W × 155 mm(6.10 in)H × 480 mm(18.90 in)D, 18 kg (634.9 oz) |
| Accessories | H.V. TEST LEAD 9615 (high voltage side and return, 1 each), Power cord (1), spare fuse (1) |